How well do clinic-based blood pressure measurements agree with the mercury standard?

Kim JW, Bosworth HB, Voils CI, Olsen M, Dudley T, Gribbin M, Adams M, Oddone EZ. How well do clinic-based blood pressure measurements agree with the mercury standard? Journal of general internal medicine. 2005 Jul 1; 20(7):647-9.

Dimensions for VA is a web-based tool available to VA staff that enables detailed searches of published research and research projects.

If you have VA-Intranet access, click here for more information vaww.hsrd.research.va.gov/dimensions/

VA staff not currently on the VA network can access Dimensions by registering for an account using their VA email address.
   Search Dimensions for VA for this citation
* Don't have VA-internal network access or a VA email address? Try searching the free-to-the-public version of Dimensions



Abstract:

BACKGROUND: Obtaining accurate blood pressure (BP) readings is a challenge faced by health professionals. Clinical trials implement strict protocols, whereas clinical practices and studies that assess quality of care utilize a less rigorous protocol for BP measurement. OBJECTIVE: To examine agreement between real-time clinic-based assessment of BP and the standard mercury assessment of BP. DESIGN: Prospective reliability study. PATIENTS: One hundred patients with an International Classification of Diseases-9th edition code for hypertension were enrolled. MEASURES: Two BP measurements were obtained with the Hawksley random-zero mercury sphygmomanometer and averaged. The clinic-based BP was extracted from the computerized medical records. RESULTS: Agreement between the mercury and clinic-based systolic blood pressure (SBP) was good, intraclass correlation coefficient (ICC) = 0.91 (95% confidence interval (CI): 0.83 to 0.94); the agreement for the mercury and clinic-based diastolic blood pressure (DBP) was satisfactory, ICC = 0.77 (95% CI: 0.62 to 0.86). Overall, clinic-based readings overestimated the mercury readings, with a mean overestimation of 8.3 mmHg for SBP and 7.1 mmHg for DBP. Based on the clinic-based measure, 21% of patients were misdiagnosed with uncontrolled hypertension. CONCLUSIONS: Health professionals should be aware of this potential difference when utilizing clinic-based BP values for making treatment decisions and/or assessing quality of care.